Series 500 Monitor

Series 500/505 Monitor



The Series 500 includes the same features and functionality of the Series 300 but with the added feature of onboard and PC data logging.




Monitor Specifications

Series 500 Monitor Base

Measurement units ppm or mg/m 3
Removable / replaceable sensor head Yes
On-board alarm Yes
Alarm status displayed Yes
Low battery indication Yes
Alarm mute function Yes
Sensor condition status Yes
Stand-by mode Yes
External signal for alarms & control Transistor output, 150 mA max
External signal functions Low Alarm, High Alarm & Control
Analogue output 0 - 5 V
Power supply 12 VDC, 800 mA
Rechargeable battery pack (optional) Standard - 9.6V Ni-MH (5 hours operation)
Portable (with optional battery pack) Yes
Permanently fixable Screw fix
Remote Sensor Capability Yes
Enclosure rating IP20 & NEMA 1 equivalent
Size (with sensor head) 195 x 122 x 54 (mm); 7? x 4¾ x 2? (in)
Weight (with sensor head and battery) < 460 g; < 16 oz
Approvals Part 15 of FCC Rules
EN 50082-1: 1997
EN50081-1: 1992
Data interface with PC Serial RS232
On Board Data Logging 8,000 data points
Data logging independent of PC Yes
Data logging direct to PC Yes
On-board real-time clock Yes
Data logging software supplied Yes

Ozone Sensor Head Specifications

Sensor Heads Low High Leak
Measurement range 0.000 - 0.500 ppm 0.50 - 20.00 ppm 0.00 - 50.00 ppm
Accuracy < 0.008 ppm (0 - 0.100 ppm)
± 10% (0.100 - 0.500 ppm)
± 10% (0.20 - 2.00 ppm)
± 15% (2.00 - 20.00 ppm)
± 20%,
15 seconds after Reset
Resolution 0.001 ppm 0.01 ppm 0.01 ppm
Controlled Atmosphere Available on Request Available on Request Available on Request
Version
T90 Response < 60 seconds (T90) < 35 seconds (T90) < 10 seconds (T90)
Sensor type Gas-sensitive semiconductor
Operating temperature range -5°C to 50°C; 23°F to 122°F (sensor head and base unit)
Relative humidity limit
95% maximum (sensor head and base unit)

VOC Sensor Head Specification*

Measurement range 0 to 400 ppm toluene
Accuracy < +/- 10 ppm (0 - 200 ppm)
< +/- 10% ( > 200 ppm)
T90 response < 60 s
Sensor type Gas-sensitive semiconductor
Operating temperature range -20°C to 60°C
Operating relative humidity range 5 to 95% non-condensating

* Specific Calibrations to other VOC's available - contact Aeroqual Sales

Ammonia Sensor Head Specification

Measurement range 0 to 1000 ppm
Accuracy < +/- 5 ppm (0 - 100 ppm)
< +/- 10% ( 100 - 1000 ppm)
T90 response < 60 s
Sensor type Gas-sensitive semiconductor
Operating temperature range -20°C to 60°C
Operating relative humidity range 5 to 95% non-condensating

CO Sensor Head Specification

Measurement range 0 to 2000 ppm carbon monoxide
Accuracy < +/- 10 ppm in the range 0 to 400 ppm
T90 response < 150 s
Sensor type Gas-sensitive semiconductor
Operating temperature range 0°C to 70°C
Operating relative humidity range 5 to 95% non-condensating
Approvals UL 2034, BS 7860

Additional Product Informationhgm300.pdf


Sensor Head Spec. sheethgm300.pdf