Series 500 Monitor
The Series 500 includes the same features and functionality of the Series 300 but with the added feature of onboard and PC data logging.
Monitor Specifications
Series 500 Monitor Base
Measurement units | ppm or mg/m 3 |
Removable / replaceable sensor head | Yes |
On-board alarm | Yes |
Alarm status displayed | Yes |
Low battery indication | Yes |
Alarm mute function | Yes |
Sensor condition status | Yes |
Stand-by mode | Yes |
External signal for alarms & control | Transistor output, 150 mA max |
External signal functions | Low Alarm, High Alarm & Control |
Analogue output | 0 - 5 V |
Power supply | 12 VDC, 800 mA |
Rechargeable battery pack (optional) | Standard - 9.6V Ni-MH (5 hours operation) |
Portable (with optional battery pack) | Yes |
Permanently fixable | Screw fix |
Remote Sensor Capability | Yes |
Enclosure rating | IP20 & NEMA 1 equivalent |
Size (with sensor head) | 195 x 122 x 54 (mm); 7? x 4¾ x 2? (in) |
Weight (with sensor head and battery) | < 460 g; < 16 oz |
Approvals | Part 15 of FCC Rules EN 50082-1: 1997 EN50081-1: 1992 |
Data interface with PC | Serial RS232 |
On Board Data Logging | 8,000 data points |
Data logging independent of PC | Yes |
Data logging direct to PC | Yes |
On-board real-time clock | Yes |
Data logging software supplied | Yes |
Ozone Sensor Head Specifications
Sensor Heads | Low | High | Leak |
Measurement range | 0.000 - 0.500 ppm | 0.50 - 20.00 ppm | 0.00 - 50.00 ppm |
Accuracy | < 0.008 ppm (0 - 0.100 ppm) ± 10% (0.100 - 0.500 ppm) |
± 10% (0.20 - 2.00 ppm) ± 15% (2.00 - 20.00 ppm) |
± 20%, 15 seconds after Reset |
Resolution | 0.001 ppm | 0.01 ppm | 0.01 ppm |
Controlled Atmosphere | Available on Request | Available on Request | Available on Request |
Version | |||
T90 Response | < 60 seconds (T90) | < 35 seconds (T90) | < 10 seconds (T90) |
Sensor type | Gas-sensitive semiconductor | ||
Operating temperature range | -5°C to 50°C; 23°F to 122°F | (sensor head and base unit) | |
Relative humidity limit | 95% maximum (sensor head and base unit) |
VOC Sensor Head Specification*
Measurement range | 0 to 400 ppm toluene |
Accuracy | < +/- 10 ppm (0 - 200 ppm) < +/- 10% ( > 200 ppm) |
T90 response | < 60 s |
Sensor type | Gas-sensitive semiconductor |
Operating temperature range | -20°C to 60°C |
Operating relative humidity range | 5 to 95% non-condensating |
* Specific Calibrations to other VOC's available - contact Aeroqual Sales
Ammonia Sensor Head Specification
Measurement range | 0 to 1000 ppm |
Accuracy | < +/- 5 ppm (0 - 100 ppm) < +/- 10% ( 100 - 1000 ppm) |
T90 response | < 60 s |
Sensor type | Gas-sensitive semiconductor |
Operating temperature range | -20°C to 60°C |
Operating relative humidity range | 5 to 95% non-condensating |
CO Sensor Head Specification
Measurement range | 0 to 2000 ppm carbon monoxide |
Accuracy | < +/- 10 ppm in the range 0 to 400 ppm |
T90 response | < 150 s |
Sensor type | Gas-sensitive semiconductor |
Operating temperature range | 0°C to 70°C |
Operating relative humidity range | 5 to 95% non-condensating |
Approvals | UL 2034, BS 7860 |
Additional Product Information