Model 3792 Facility Monitoring Laser Particle Sensor
Cleanroom Monitoring - Pharmaceutical, Aerospace/ Defense, Electronics, MEMS, Semiconductor
Food Processing
Medical / Hospitals
Features
- 0.2um/0.3um sensitivity
- Built-in sonic nozzle for accurate and consistent flow
- Built-in LEDs for sensor status at a glance
- Compact stainless body
- Temperature/Relative Humidity sensors available as option
- Status LED shows Particle, LD Status, and Flow
Specifications
Model | 3792-01 | 3792-03 |
Sampling Air Flow | 2.83 l/min | 1 l/min |
Sensitivity | 0.2 µm, 0.3 µm | 0.2 µm, 0.3 µm |
Condensation Range | 1,000,000 pieces/cf | 3,000,000 pieces/cf |
Output | Type A | RS-485 Output Connector Type RS-485 IN Modular Connector 8-pin (rear panel) RS-485 OUT Modular Connector 8-pin(rear panel) Transmission Setting: DIP switch 8-pin(front panel) Address Setting(0-999): Rotary DIP switch, 10 positions x 3 (front panel) Multiplexer Output Connector Type: BNC Connector |
Type B | 4-20mAOutput
Output: Data for 2 particle size ranges and Sensor Status Connector Type: D-sub Connector 9-pin (rear panel) Indicative Resolution: 1/1000 of the full scale Full Scale: 3 levels (100, 10,000/1,000,000 pieces / cf) Full Scale Selection: Sliding DIP switch 8-pin(front panel) Multiplexer Output Connector type: BNC Connector |
Display | 5-Digit LCD Optional Display unit available (needs to be exchanged with the supporting casing) Condensation Display (Running Average for pre-set time) SIZE key gives selection of Particle size range to monitor |
Status LED | PARTICLE: Turn Red when picking the signal of the particle
STATUS: Keep Green when Optics are operating normally. FLOW: Turn Green when Sampling Flow rate is normal |
Power Supply | AC 100V/240V |
Dimensions | 4.6" x 5.3" x 6.7" (118×135×170 mm) (Excluding sampling air nozzle on the upper surface and other interface outfit. Also excludes rubber feet.) |
Weight | Approx. 6.6lbs (Approx. 3kg) |